====== SOP ====== == System Power On == - Press “system ON” on Siemens switch box (mounted on wall) * Do NOT press power button on the Syngo computer (HP Z2 computer) - Click “System Check” => wait until all indicators on the “System Status Overview” are green - Troubleshooting: If Syngo computer does not automatically turn on or some indicators are red * TerraX was not shut down properly * Follow System Power Off procedures below and repeat power on procedures == System Power Off == - Ordered List ItemClick power icon on home screen => shutdown console => wait until Syngo computer turns off - Press “system OFF” on Siemens switch box (mounted on wall)