SOP

System Power On
  1. Press “system ON” on Siemens switch box (mounted on wall)
    • Do NOT press power button on the Syngo computer (HP Z2 computer)
  2. Click “System Check” ⇒ wait until all indicators on the “System Status Overview” are green
  3. Troubleshooting: If Syngo computer does not automatically turn on or some indicators are red
    • TerraX was not shut down properly
    • Follow System Power Off procedures below and repeat power on procedures
System Power Off
  1. Ordered List ItemClick power icon on home screen ⇒ shutdown console ⇒ wait until Syngo computer turns off
  2. Press “system OFF” on Siemens switch box (mounted on wall)